Does Exercise Intensity Influence Wrist-Worn Heart Rate Accuracy? Validation of the Apple Watch Following Treadmill Walking and Running in Healthy Young Adult Males
Abstract
Objective: Wrist-worn wearable technologies are widely used for heart rate (HR) monitoring in both fitness and research settings. However, their measurement performance may vary according to exercise intensity, and evidence regarding the consistency of this intensity dependence remains limited. This study evaluated the agreement of post-exercise HR measurements obtained from the Apple Watch Series 8 compared with those obtained from a fingertip pulse oximeter used in clinical practice following standardized 6-minute treadmill exercise performed at three intensities in healthy young adult males.
Materials and Methods: This method comparison study enrolled fifteen healthy young adult males who completed three standardized 6-minute treadmill exercise bouts of walking (4.0 km/h), brisk walking (7.2 km/h), and running (9.0 km/h). Immediately after each bout, HR was measured using both devices. Agreement was assessed using Pearson’s correlation coefficient (r), Lin’s concordance correlation coefficient (CCC), Bland-Altman analysis, and absolute error metrics.
Results: Good-to-excellent correlations were observed across all intensities (r = 0.981-0.993), with high concordance (CCC = 0.95-0.99). Bland-Altman analysis demonstrated small systematic bias (within ±2 bpm) and narrow limits of agreement. Absolute error remained low across conditions (MAE: 1.3-1.5 bpm; MAPE < 2%). Although overall agreement was high, concordance was slightly reduced during running compared with lower intensities.
Conclusion: The Apple Watch Series 8 provided accurate post-exercise HR measurements across walking, brisk walking, and running conditions, with a modest intensity-dependent decline in agreement at higher workloads.
Keywords
Full Text:
PDF (Early Version)References
Achten J, Jeukendrup AE. Heart rate monitoring: applications and limitations. Sports Med 2003;33(7):517-38.
Halson SL. Monitoring training load to understand fatigue in athletes. Sports Med 2014;44(Suppl 2):139-47.
Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med 2020;3:18.
Fuller D, Colwell E, Low J, Orychock K, Tobin MA, Simango B, et al. Reliability and validity of commercially available wearable devices for measuring steps, energy expenditure, and heart rate: systematic review. JMIR Mhealth Uhealth 2020;8(9):e18694.
Chan ED, Chan MM, Chan MM. Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations. Respir Med 2013;107(6):789-99.
Dehghanojamahalleh S, Kaya M. Sex-related differences in photoplethysmography signals measured from finger and toe. IEEE J Transl Eng Health Med 2019;7:1-7.
Orini M, Guvensen G, Jamieson A, Chaturvedi N, Hughes AD. Movement, sweating, and contact pressure as sources of heart rate inaccuracy in wearable devices. In: 2022 Computing in Cardiology (CinC). IEEE; 2022.
Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med 2017;7(2):3.
Hough P, Glaister M, Pledger A. The accuracy of wrist-worn heart rate monitors across a range of exercise intensities. J Phys Act Res 2017;2(2):112-6.
Gillinov S, Etiwy M, Wang R, Blackburn G, Phelan D, Gillinov AM, et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med Sci Sports Exerc 2017;49(8):1697-703.
Ho WT, Yang YJ, Li TC. Accuracy of wrist-worn wearable devices for determining exercise intensity. Digit Health 2022;8:20552076221124393.
Zhang Y, Weaver RG, Armstrong B, Burkart S, Zhang S, Beets MW. Validity of wrist-worn photoplethysmography devices to measure heart rate: a systematic review and meta-analysis. J Sports Sci 2020;38(17):2021-34.
ATS Committee on Proficiency Standards for Clinical Pulmonary Function Laboratories. ATS statement: guidelines for the six-minute walk test. Am J Respir Crit Care Med 2002;166(1):111-7.
Du H, Wonggom P, Tongpeth J, Clark RA. Six-minute walk test for assessing physical functional capacity in chronic heart failure. Curr Heart Fail Rep 2017;14(3):158-66.
Pasadyn SR, Soudan M, Gillinov M, Houghtaling P, Phelan D, Gillinov N, et al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther 2019;9(4):379-85.
Craig CL, Marshall AL, Sjostrom M, Bauman AE, Booth ML, Ainsworth BE, et al. International physical activity questionnaire: 12-country reliability and validity. Med Sci Sports Exerc 2003;35(8):1381-95.
Rogers DW, Himariotis AT, Sherriff TJ, Proulx QJ, Duong MT, Noel SE, et al. Test-retest reliability and concurrent validity of photoplethysmography finger sensor to collect measures of heart rate variability. Sports (Basel) 2025;13(2):29.
Cozgarea A, Cozma D, Teodoru M, Lazar-Hocher AI, Cirin L, Faur-Grigori AA, et al. Heart rate recovery: up to date in heart failure-a literature review. J Clin Med 2024;13(11):3328.
Ozemek C, Bonikowske A, Christle J, Gallo P. ACSM's guidelines for exercise testing and prescription. Lippincott Williams & Wilkins; 2025.
Refbacks
- There are currently no refbacks.

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.